Product Overview
E+H Memosens CCS51D Free Chlorine Sensor
Memosens CCS51D
The Endress+Hauser Memosens CCS51D is a high-precision digital sensor designed for continuous free chlorine measurement. Featuring a membrane-covered amperometric 2-electrode measuring cell and Memosens contactless technology, it provides long-term stability and fast response times for efficient process control.
High-Precision Free Chlorine Measurement
The Endress+Hauser Memosens CCS51D is a robust, low-maintenance amperometric sensor specifically engineered for real-time monitoring of free chlorine in water treatment and process applications. Operating with a membrane-covered 2-electrode measuring cell, the sensor reduces hypochlorous acid (HOCl) at the cathode to deliver linear, accurate, and repeatable disinfectant readings. Its fast response time (t90 < 25 s) allows operators to respond promptly to process fluctuations, ensuring safe water disinfection and preventing over-dosing.
Durable Membrane Design and Flexible Range
Built with a resistant POM shaft, PVDF membrane cap, and FKM seal rings, the CCS51D withstands aggressive water conditioning environments and process stresses. The sensor is available in versatile measuring ranges—from trace detection (0 to 5 mg/L) up to standard (0 to 20 mg/L) and high-concentration ranges (0 to 200 mg/L)—making it suitable for drinking water safety, swimming pools, and beverage bottling loops.
Contactless Memosens Digital Technology
Integrating Memosens digital communication technology, the CCS51D eliminates signal interference from moisture and electromagnetic noise. The sensor stores calibration and process data directly within the sensor head, allowing laboratory pre-calibration and seamless hot-swapping in field installations. Connected to Liquiline multiparameter transmitters, it integrates easily with pH and ORP sensors for comprehensive liquid analysis control.
Applications
-
Drinking Water Treatment: Ensures safe disinfection monitoring and precise chlorine dosing in municipal supply networks.
-
Process Water & Utilities: Detects the presence or absence of free chlorine in cooling towers, washing systems, and utility lines.
-
Food & Beverage Industry: Guarantees hygienic rinsing water and packaging sterilization for food safety compliance.
-
Pool & Leisure Water: Provides automated disinfectant dosing for swimming pool water hygiene and safe bathing conditions.
Specifications
-
Measurement Principle: Amperometric 2-electrode measuring cell with PVDF membrane.
-
Measured Parameters: Free chlorine / Hypochlorous acid (HOCl).
-
Measurement Ranges: Trace (0 to 5 mg/L), Standard (0 to 20 mg/L), High (0 to 200 mg/L) HOCl.
-
Response Time: t90 < 25 s for rapid process control.
-
Operating Temperature: 0 to 55°C (32 to 130°F), non-freezing.
-
Operating Pressure: Max. 1 bar (14.5 psi).
-
Temperature Sensor: Integrated NTC 10k temperature sensor for automatic compensation.
-
Materials: POM sensor shaft, PVDF membrane cap, FKM sealing ring.
-
Digital Protocol: Contactless inductive head with Memosens protocol.
FAQ
-
Q: What chlorine species does the Memosens CCS51D measure? A: It measures free chlorine, specifically hypochlorous acid (HOCl) dissolved in water.
-
Q: Is the CCS51D compatible with Liquiline transmitters? A: Yes, it connects via Memosens digital inductive head directly to Liquiline multiparameter transmitters.
-
Q: Can the sensor be pre-calibrated before field installation? A: Yes, Memosens technology stores calibration data in the sensor head, enabling pre-calibration in the lab and quick field plug-and-play swap.