Designed for high-accuracy mass flow and density measurement of liquids and gases across demanding industrial processes. Features a robust remote transmitter design with multivariable capabilities and advanced digital communication.
Experience unparalleled measurement accuracy and long-term operational reliability with the Endress+Hauser Proline Promass F 500 Coriolis Flowmeter. Built on a sophisticated bent dual-tube measuring system, this state-of-the-art instrument delivers exceptional performance for both liquid and gas applications under demanding process conditions. Whether operating in chemical processing, oil and gas, energy, or general industrial automation, the Promass F 500 provides robust, drift-free measurement independent of fluid properties like viscosity or density.
Advanced Architecture and Robust Design The Promass F 500 utilizes a remote transmitter design, housing the electronics separately from the sensor to withstand extreme environmental challenges, such as heavy pipeline vibrations, high ambient temperatures, or underground installations. Available in rugged aluminum or cast stainless steel housings, the measuring system guarantees IP66/67 (Type 4X) protection. It integrates seamlessly into modern industrial networks via Modbus TCP over Ethernet-APL or Fast Ethernet, offering flexible I/O configurations including current outputs, pulse/frequency/switch outputs, and relay outputs.
Multivariable Measurement and Smart Capabilities Beyond precise mass and volume flow measurement, the device continuously tracks fluid density and process temperature. Integrated Heartbeat Technology enables comprehensive in-situ verification, self-diagnostics, and health monitoring without interrupting the live process, significantly reducing maintenance overhead and ensuring regulatory compliance. The intuitive local graphic display with touch control, web server access, and the SmartBlue app make configuration, data management, and commissioning fast and secure. Trust this premier Coriolis flowmeter to optimize your process control, maximize safety, and deliver repeatable accuracy batch after batch.
Application Areas
-
Chemical & Petrochemical Industry: Measurement of corrosive, aggressive, or high-viscosity liquids, acids, solvents, and liquefied gases.
-
Oil & Gas Sector: Custody transfer, refined products measurement, mass flow tracking, and gas fraction handling (GFH) for two-phase media.
-
Food & Beverage / Life Sciences: Hygienic processes requiring strict cleanability, robust stainless steel wetted materials, and high-precision dosing.
-
Energy & Power Utilities: Boiler feed water monitoring, fuel gas measurement, and energy management systems.
Technical Specifications
-
Measuring Principle: Coriolis mass flow measurement
-
Nominal Diameter Range: DN 8 to 250 (3/8 to 10 inch)
-
Maximum Mass Flow Range (Liquids): Up to 2,200,000 kg/h (80,850 lb/min)
-
Base Accuracy (Mass Flow – Liquids): ±0.05% o.r. (PremiumCal) or ±0.10% o.r. (Standard)
-
Medium Temperature Range: -196 to +350 °C (-320 to +662 °F)
-
Communication Protocols: Modbus TCP (via Ethernet-APL or 100 Mbit/s Fast Ethernet), optional HART, PROFIBUS, PROFINET, EtherNet/IP
-
Power Supply: DC 24V or AC 100 to 240V
-
Degree of Protection: IP66/67, Type 4X enclosure
-
Wetted Materials: Stainless steel (1.4404/316L, 1.4539/904L) or Alloy C22 (2.4602)
FAQ
-
Q: Can the Promass F 500 handle two-phase or gassy liquids?
-
A: Yes. It features an integrated Gas Fraction Handler (GFH) function that stabilizes flow and density measurements in the presence of entrained gas bubbles or micro-bubbles.
-
-
Q: How is remote configuration and data access handled?
-
A: The device supports configuration via its backlit touch display, an integrated web server accessed through an Ethernet browser, SmartBlue app via optional WLAN, or asset management tools like FieldCare and DeviceCare.
-
-
Q: Is a zero point adjustment required after installation?
-
A: All devices are factory-calibrated under reference conditions. Field zero adjustment is generally unnecessary unless targeting maximum accuracy at extremely low flow rates or under extreme process conditions.
-